C. Please inspect the hardware, if the current value is inconsistent
with the value in the tolerance error table after the value of
calibration function is transformed to actual current value.
DS 3640
Function
Setting
Range of Transformation Value
Lo
0.1A
0 ~ 0.5 A
MIDL
13A
11 ~ 14 A
MIDH
26A
23 ~ 27 A
Hi
38A
34 ~ 39 A
DS 6024
Function
Setting
Range of Transformation Value
Lo
0.06A
0 ~ 0.5 A
MIDL
7.8A
6.6 ~ 9 A
MIDH
15.6A
14 ~ 17.2 A
Hi
22.8A
21 ~ 24.6 A
DS 8018
Function
Setting
Range of Transformation Value
Lo
0.045A
0 ~ 0.1 A
MIDL
5.9A
5.3 ~ 6.5 A
MIDH
12A
10.8 ~ 13.2 A
Hi
17A
15.3 ~ 18.7 A