After receiving the control order from the Master, all Slaves will be locked on
the SLAVE page that all keys except
are disabled so that slaves can not
be operated by press keys except controlled by the Master as graphically
demonstrated below.
:
:
L
L
L
Once the Master and Slave settings are done, users can then operate the Master
only to set up the output voltage and current.
To release from this operation mode (parallel operation mode in this example),
press
(LCL) to access to the series/parallel setting page and turning the
knob to select OFF for series/parallel selection to leave the series or parallel
operation mode and back to the local operation mode that the
DS3640/DS6024/DS8018/DS10014 can be controlled by press the keys again.
It shall not change the operation mode while DS3640/DS6024/DS8018/DS10014
is in output state or it will cause the Master’s communication failure and
display error message.