23
2.2.3 Open and Short Calibration
The MT4080 provides open/short calibration capability so the
user can get better accuracy in measuring high and low impedance.
We recommend that the user performs open/short calibration if the
test level or frequency have been changed.
Open Calibration
First, remaining the measurement terminals at the open status.
Press the CAL key shortly (no more than two second), the
LCD will display:
This calibration takes about 10 seconds. After it is finished, the
MT4080 will beep to show that the calibration is done.
Short Calibration
To perform the short calibration, insert the Shorting Bar into
the measurement terminals. Press the CAL key for more than
two second, the LCD will display:
This calibration takes about 10 seconds. After it is finished, the
MT4080 will beep to show that the calibration is done.