24
2.2.4 Display
Speed
The MT4080 provide two different display speeds (Fast/Slow). It
is controlled by the Speed key. When the speed is set to fast, the
display will update 4.5 readings every second. When the speed is
set to slow, it’s only 2.5 readings per second.
2.2.5 Relative
Mode
The relative mode lets the user to make quick sort of a bunch of
components. First, insert the standard value component to get the
standard value reading. (Approximately 5 seconds in Fast Mode to
get a stable reading.) Then, press the Relative key, the primary
display will reset to zero. Remove the standard value component
and insert the unknown component, the LCD will show the value
that is the difference between the standard value and unknown
value.
2.2.6 Range
Hold
To set the range hold, insert a standard component in that
measurement range. (Approximately 5 seconds in Fast Mode to get
a stable reading.) Then, by pressing the Range Hold key it will
hold the range within 0.5 to 2 times of the current measurement
range. When the Range Hold is press the LCD display: