V. Photomicrographic Procedure (Only on AE2000 TRI)
y
To ensure vibration free operation, set the microscope on a sturdy vibration free table or a bench
with a vibration proof device.
y
Turn the beam splitter knob of the eyepiece tube to the “photo position” (Fig.31). The photo exit is
activated, 80% of the light will enter the camera.
y
Select a blue filter for routine application. An additional colour-compensating filter can also be used
depending on the colour rendition.
y
A change of depth of focus, contrast and resolution of image is attainable with an aperture setting of
about 2/3 maximum. Fine setting of the condenser aperture is depending on the individual sample.
y
For photomicrographic procedures, refer to the manual of the specific camera being used.
Fig.31