11.2 Detailed Assembly Procedures
11.2.1 Installing the Bulb
WARNING
The bulb, the lamp socket and areas near these will be extremely hot during and right after use. Set the
main switch off; disconnect the power cord from the wall outlet, then wait until lamp housing has cooled
down before replacing the bulb with a new of the designated type.
●
Use only the designated bulb 12V100W Halogen lamp. (PHILIPS 7724)
●
Fully loosen the lamp housing clamping screw
of the lamp housing cover with provided Allen
screwdriver.
●
Lift the lamp housing cover
upward to remove.
●
Holding the bulb with gloves or a piece of gauze, depress the bulb clamping lever
and insert bulb
pin fully into the pin holes
on the lamp socket. Gently release the bulb clamping lever to the
original position to secure the bulb.
●
Side the lamp housing cover onto the housing base from the above. Tighten the clamping screw
while pressing downward on the cover.
Fig. 21
Fig. 22
CAUTION
To prevent reduced bulb life or cracking, do not touch the bulb with bare hands. If fingerprints are accidently left on the
bulb, wipe the bulb with a soft cloth.
NOTE
When you dispose the device and accessory kit, comply with related law and recycling plan. Disposal of light bulb can
cause the environmental pollution.
When you dispose the packing material, comply with related law and recycling plan.