Troubleshooting: 
 
ELECTRICAL PROBLEMS 

 
 

PROBLEM CAUSE 

SOLUTION 

The bulb does not work.  Outlet inoperative.  

 
Cable not connected.  
 
Bulb burned out. 
Fuse blown. 
Wrong Bulb. 

Have it repaired by a qualified 
specialised technician. 
Connect the cable to the power 
source.  
Replace bulb. 
Replace fuse. 
Replace with appropriate bulb. 

Bulb burns out 
immediately. 

Wrong bulb. 

Replace with the appropriate bulb. 

Bulb flickers. 

The bulb is not inserted correctly into 
socket. 
Bulb on the point of burning out. 
Fuse cover badly closed. 
Bad connection with power source. 
 

Insert bulb correctly. 
 
Replace bulb. 
Close correctly. 
Have repaired by a qualified 
specialised technician. 

Fuse blown quickly 

Wrong fuse. 

Replace with appropriate fuse. 

Fuse blown instantly 

Short circuit. 

Have repaired by a qualified 
specialised technician. 

 
 

IMAGE QUALITY 

 
 

PROBLEM CAUSE 

SOLUTION 

No image. 

Objectives badly positioned. 

Adjust, until end of the dovetail. 

Poor resolution. 

Eyepieces dirty. 
Objectives dirty. 

Clean eyepieces. 
Clean objectives. 

Spots, or stains in field 
of view. 

Eyepieces dirty. 

Clean eyepieces. 

* NB. Stains in field of view may also be attributed to dirt on the inside of the eyepiece. It is recommended 
therefore that the lens be cleaned by a recognised service technician.  
 
 

MECHANICAL PROBLEMS 
 
 
PROBLEM CAUSE 

SOLUTION 

It does not stay in focus.   The head drops down. 

Adjust the tension of the coarse focus 
knob. 

 

 
 
 

 
 
 
 
 
 

 
 

 

-8- 

 

Summary of Contents for PS10100001

Page 1: ...Instruction Manual S 10 20 Series Copyright 02 02 Motic Microscopes European Division...

Page 2: ...iece holder tube 4 Head 15 Incident illumination switch I 16 Transmitted illumination switch T 12 Focussing knob 10 Diopter corrector 1 Eyepiece 17 Base 11 Head locking screw 5 Objectives 6 Incident i...

Page 3: ...e it for long periods of time or should it become necessary to transport it to a technical service for any repair or maintenance procedure The box should contain the following components depending on...

Page 4: ...ss for the observation of slides or thin or transparent samples such as leaves insect wings etc The other is opaque black and white and can be used for non transparent samples or for dissection The be...

Page 5: ...cessary to repeat these steps every time the object is changed so long as the objects are of similar thickness or height 4 Adjust again focus knob until the image is completely detailed D Adjusting th...

Page 6: ...trical maintenance Except S 10 P 1 Changing the transmitted illumination bulb Only for S 20 2LOO a Loosen the stage support screw Fig 1 with the key provided and remove the stage 9 b Carefully remove...

Page 7: ...ance 1 Adjusting focus tension Tension comes pre adjusted by the manufacturer The optimum tension point is that which allows as smooth a movement possible of the focus knobs but does not allow the sta...

Page 8: ...orrectly Have repaired by a qualified specialised technician Fuse blown quickly Wrong fuse Replace with appropriate fuse Fuse blown instantly Short circuit Have repaired by a qualified specialised tec...

Page 9: ...of the required revision Warrantee All MOTIC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through misuse...

Reviews: