B. Mechanical 

Maintenance 

 

4.  Adjusting the tension of the coarse focus knob.  

 

The collar to adjust the tension of the coarse focus knob (Fig. 1) is situated between 
the coarse focus knob (9) and the arm (12).  
The tension of the coarse focus is pre adjusted by the manufacturer. The optimum 
tension point is that which permits the coarse focus knob to move as smoothly as 
possible, without the head (7) sliding down by itself. 

 

a.  To adjust tension first loosen the screw situated in one of the four orifices of the 

collar with a 2mm hexagonal key.  

 

b.  To tighten the tension of the coarse focus knob, turn the collar in the clockwise. To 

loosen it, turn anti clockwise.  

 

c. Tighten 

hexagonal 

screw. 

 

5.  Adjusting the rack stop. 

 

The 40X objective use a retractable 
security system to avoid damage to the 
slide or to the front of the lens should they 
come into contact. An additional security 
measure is the security rack stop screw 
that regulates the descending movement 
of the head. It is adjusted by the 
manufacturer for the observation of 
standard slides with a cover slip 0.17mm 
thick; but for the observation of other types 
of slides, readjustment may be necessary.   

 
 

Tensión 

adjustment

collar

Set screw

Screw used to 

adjust the height 

of the head

a.  Firstly, loosen the rack stop 

screw. (Fig. 1) 

 
b.  Next, unscrew the adjustment 

screw. (Fig. 1) 

 

c.  Turn the fine focus knob (10) 

until at mid point (not applicable 
for SFC-3A). 

 

Fig. 1 

d.  Focus on the slide only using the coarse focus control (9) with the 4X objective (5) 

firstly, and then the 10X. 

 

e.  Tighten the adjustment screw only until firm, so that the head (3) cannot descend.  

 

f. 

Tighten the rack stop screw. 

 
 
 
 
 
 
 
 
 
 
 