background image

6.4A: EXAMPLE: Programming Memory 

Program the following operation sequences to practice saving them with your Duo.

   

Operating

Mode   

Massage

  Mode   

Massage

  Mode   

Expression

  Mode   

Expression

   Mode   

Speed / Vacuum

Level  

Level 4

Level 5

Level 3

Level 4

Operating

Time  

1 Minnute

30 Seconds

1 Minute

Until Pumping

Session Completed

1. Check the LCD display to see if there is an existing saved operating 

sequence, the Memory Icon will be displayed if there is an existing memory 

sequence saved. 

2. If there is a saved sequence, delete it by pressing and holding MODE 

button for 3 seconds. The Memory Icon will disappear once the sequence 

has been deleted.

3. Set the Massage Mode on level 4 and Expression Mode on level 3.

4. Change the Operating Mode to Massage Mode on level 4.

5. Begin the memory function by pressing and holding the POWER button for 

3 seconds. The Memory Icon will flash to indicate the sequence is saving. 

6. Press the MODE button until the Massage Mode icon is illuminated, then 

adjust Speed/Vacuum to Level 4 by pressing the INCREASE or DECREASE 

buttons. Then allow the pump to operate for 1 minute. 

7. Increase the Speed/Vacuum to level 5 by pressing the INCREASE button, 

and let the pump run for 30 seconds.

8. Change to Expression Mode by pressing the MODE button and the Expression 

Mode Icon will be displayed. Adjust Speed/Vacuum to level 3 by pressing the 

INCREASE or DECREASE buttons. Allow the pump to run for 1 minute.

9. Press the INCREASE button briefly to increase Speed/Vacuum to level 4, 

and then press and hold the POWER button for 3 seconds in order to save 

the operating sequence.

10. The last programmed step will continue to operate until stopped by the 

user or until the timer reaches 30 minutes.

11. To run the saved operating sequence, press and hold the DECREASE 

button for 3 seconds.

12. To stop running the saved operating sequence, press and hold the 

INCREASE button for 3 seconds. The pump will return to your previous settings.

6.4B Frequency Function

This function is available during expression mode, and effects the number of 

times your pump pulls per minute.

• There are 3 frequency levels (1-3), that you can select during 

Expression Modes by pressing and holding the INCREASE or DECREASE 

buttons for 1 second. 

• The LCD Display will show [CY: --], to indicate the current frequency level. 

• Adjust the frequency level by pressing either the INCREASE or 

DECREASE buttons.

• The frequency will be displayed for 4 seconds, then the pump will 

automatically return to normal operating mode.

Note: Increasing the frequency may also increase the noise and vibration. 

7. Preparing to Pump

1. Thoroughly wash your hands with soap and water before touching any 

pump parts.

2. Check each component for wear or damage before use and 

replace parts when/if necessary.

3. Make sure that all components that come into contact with breast milk 

are properly sanitized. 

4. Assemble the breast shield, diaphragm, and bottle according to the 

instructions and connect to the pump body via the tubing.

5. Ensure tubing is tightly connected to prevent air leakage.

6. Check battery level before use to ensure the device is fully charged.

7. Massage the breast for about 5 minutes with a warm towel to 

promote expression.

Summary of Contents for Motif Duo

Page 1: ...the Motif Duo Double Electric Breast Pump Motif Duo Operation Manual Motif Medical 8 The Green Suite 5601 Dover DE 19901 Phone 844 272 8390 Email info motifmedical com Web www motifmedical com Duo EN...

Page 2: ...ram 6 Operation and Programming 6 1 LCD Display and Buttons 6 1A LCD Display Table 6 2 Powering Your Breast Pump 6 3 Operating Your Breast Pump 6 4 Programming Your Breast Pump 6 4A Memory Function 6...

Page 3: ...anty Date of Purchase 2 Important Safeguards 2 1 Intended Use The Motif Duo Double Electric Breast Pump is intended for use by lactating women to express and collect breast milk The Duo can be used as...

Page 4: ...er Submerge any pump parts that come in direct contact with breast milk 3 Bring water to a boil and allow parts to sanitize for 5 minutes 4 Once water has cooled remove parts with caution as they migh...

Page 5: ...ns MEMORY TIMER Indicates a program is stored EXPRESSION MODE POWER Displays when plugged into power adapter Switches between massage mode and expression mode Displays while in massage mode Decreases...

Page 6: ...r Breast Pump The Motif Duo software has three major components Massage and Expression Modes Frequency Function and Programmable Memory Buttons Each button has three functions Pressing the button will...

Page 7: ...response Mode Button Switches between the Massage and Expression Modes This will delete a saved memory sequence The Memory icon will turn off once the sequence has been deleted After a sequence is de...

Page 8: ...4 and then press and hold the POWER button for 3 seconds in order to save the operating sequence 10 The last programmed step will continue to operate until stopped by the user or until the timer reac...

Page 9: ...ssemble the parts Make sure the tubing is free of twists and kinks While single pumping make sure the Auxiliary Port is blocked by the Port Cover Double Check connections between the Diaphragm Assembl...

Page 10: ...hawing and Warming Milk Clearly label milk containers with the date the milk was expressed Use oldest stored milk first Thaw the bottle or bag of frozen milk by putting it in the refrigerator overnigh...

Reviews: