- 25 -
Tilt DLO
(Angle measured from vertical)
max 30°
(Tipsy Sensor)
Important!
Ensure base is on a level surface when adjusting limits
max 25°
(No Reduced Drive)
Front
Recline DLO
4.0
4.0
ELECTRONICS
ELECTRONICS
NOTE:
For detailed programming and operating instructions for ourTipsy Angle Sensors please refer
to the applicable instruction manual provided separately (
TRD0275- M11-R Remote Tipsy Sensor
Instructions
or
TRD0323- Remote Attendant Control User Manual
).
4.3 Safety Lockout and Limit Switches
WARNING! Risk of severe injury or death
• The angle at which the limit switches/lockouts are set is
critical
to the safe operation of the power
positioning system.
• Motion Concepts will not be liable for any injuries or damage sustained when adjustments are made
beyond the factory recommended settings.
• To ensure proper set-up, adjustments to safety lockouts and limits should only be performed by a
Qualified Technician.
WARNING! Never exceed the maximum recommended limits
Safety Lockouts and Limit Switches should be set up to best meet the needs of the user without
compromising the overall stability of the wheelchair.
NOTE:
Additional safety limits and lockout switches may be required for more complicated/specialized
seating systems. For information on limits/lockouts that are not identified in this manual, please contact
our Technical Service Department for assistance.
WARNING!
Following any limit or lockout adjustments,
always
test the seating system over the full
range of motion (i.e.; tilt, recline, elevate) to verify the revised set-up is functioning properly and
ensure that there are no resulting stability or interference issues.
4.3.1 Tilt/Recline Drive Lockout (DLO) Limit
All tilt or recline power positioning systems are equipped with a
Drive Lockout (DLO
) Limit to prevent the
wheelchair from being driven when the seating system is tilted or reclined beyond a pre-determined safe angle.
Depending on the type of power positioning system (PPS), the DLO limit may be configured using a Ball Switch,
or via our remote Tipsy Angle Sensor
(see section 4.3.6).
Note:
Systems equipped with a Tipsy Angle Sensor offer a reduced speed drive setting that will trigger the
wheelchair base to drive at approx. 25% of its maximum speed when the back angle extends beyond the home
position (Reduced Drive is typically set to engage at 20° and will stop when the Max DLO limit is reached)
Tilt and/or Recline Systems (Ball Switch- No Reduced Drive):
Max. DLO = 25°
Tilt and/or Recline Systems (Tipsy Sensor- Reduced Drive):
Max. DLO = 30°