- 7 -
If you have any questions or concerns regarding the instructions contained in this manual, please contact
our
Technical Service Department
for assistance.
USA
888-433-6818
CAN
800-680-4191
IMPORTANT!
RECOMMENDED TIPSY ANGLES (SAFETY LIMITS)
The tipsy angle sensors allows for the individual programming of up to 4 limits:
Home Position, Reduced
Drive Speed (RDS), Drive Lockout (DLO) Limit
and the
Max Tilt (Tilt/Recline) Limit.
HOME POSITION
(for Stander (MPSS))
Home Setting = 0° - 10° Max
HOME POSITION
(excluding Stander)
Home Setting = 0° - 15° Max
REDUCED DRIVE SPEED (RDS) LIMIT RDS Limit = 10° - 20° Max
DRIVE LOCKOUT (DLO) LIMIT
1
Max. DLO Limit = 30°
(with Reduced Drive Speed)
Max. DLO Limit = 25°
(without Reduced Drive Speed)
MAXIMUM BACK ANGLE: Max. Tilt Limit = 45°
(45T12 module)
Max. Tilt Limit = 45°
(45PAT module, Upfront, Stander)
Max. Tilt Limit = 50°
(50T and HD50T module)
Max. Tilt/Recline Limit = up to 168°
1
Note
: DO NOT exceed the maximum recommended DLO Limit.
2.0 PROGRAM MODE for the M290 REMOTE
2.2 RECOMMENDED SAFETY LIMITS
WARNING!
Risk of system damage, serious injury or death
Your power positioning system should
never
exceed the recommended safety limits. These Safety Limits are
controlled by a
Tipsy Angle Sensor
that is mounted in a fixed position on our Motion Concepts Power
Positioning Systems. The angle at which the limits are set is critical to the safe operation of the power
positioning system (refer to the recommended Safety Limts/Angles provided below).
Motion Concepts will not be liable for any injuries or damage sustained when adjustments are made beyond
the maximum recommended safety limits. If at any time your power positioning system/wheelchair is operating
beyond the recommended safety limits, or if you feel unstable at any time while operating your power positioning
system, return the system to its stable (home) position and contact your Dealer or Service Provider immediately
for assistance and/or inspection.
TIPSY ANGLE SENSOR
The
M11S Tipsy Angle Sensor
connects directly into our M-Series Seat Control, and the limits/angles can be set
using the M290 Remote Attendant Control.
Set-up or adjustment of the safety limits in the Tipsy Angle Sensor
(using the M290 Remote ) may only be performed by a Qualified Technician. (Access is password protected).