IMPORTANT! 

RECOMMENDED TIPSY ANGLES (SAFETY LIMITS)

The tipsy angle sensors allows for the individual programming of up to 4 limits:

Home Position, Reduced

Drive Speed (RDS), Drive Lockout (DLO) Limit 

and the 

Max Tilt (Tilt/Recline) Limit.

     HOME POSITION  

(for Stander (MPSS))

      Home Setting =  0° - 10° Max

        HOME POSITION

(excluding Stander)

           Home Setting =  0° - 15° Max

    REDUCED DRIVE SPEED (RDS) LIMIT       RDS Limit =  10° - 20° Max

     DRIVE LOCKOUT (DLO) LIMIT

1

                Max. DLO Limit = 30° 

(with Reduced Drive Speed)

                                                                      Max. DLO Limit = 25°  

(without Reduced Drive Speed)

    MAXIMUM BACK ANGLE:                        Max. Tilt Limit = 45°

(45T12 module)

                                                                           

Max. Tilt Limit = 45°

(45PAT module, Upfront, Stander)

                                                                                 

Max. Tilt Limit = 50°

(50T and HD50T module)

                                                                           

Max. Tilt/Recline Limit = up to 168° 

1

Note

: DO NOT exceed the maximum recommended DLO Limit.

2.0  PROGRAM MODE for the M290 REMOTE

2.2  RECOMMENDED SAFETY LIMITS 

WARNING! 

Risk of system damage, serious injury or death

Your power positioning system should 

never 

exceed the recommended safety limits. These Safety Limits are

controlled by a 

Tipsy Angle Sensor

that is mounted in a fixed position on our Motion Concepts Power

Positioning Systems.  The angle at which the limits are set is critical to the safe operation of the power
positioning system (refer to the recommended Safety Limts/Angles provided below).  

Motion Concepts will not be liable for any injuries or damage sustained when adjustments are made beyond
the maximum recommended safety limits.   If at any time your power positioning system/wheelchair is operating
beyond the recommended safety limits, or if you feel unstable at any time while operating your power positioning
system, return the system to its stable (home) position and contact your Dealer or Service Provider immediately
for assistance and/or inspection.

TIPSY ANGLE SENSOR

The 

M11S Tipsy Angle Sensor 

connects directly into our M-Series Seat Control, and the limits/angles can be set

using the M290 Remote Attendant Control. 

Set-up or adjustment of the safety limits in the Tipsy Angle Sensor

(using the M290 Remote ) may only be performed by a Qualified Technician. (Access is password protected). 

Summary of Contents for M290

Page 1: ...ational instructions specific to the Motion Concepts M290 Remote Attendant Control box This Manual MUST be given to the End User of the product USER Before using this product please read through the instructions carefully and save for future reference M290 REMOTE ATTENDANT CONTROL ...

Page 2: ...ion available at the time of publication We reserve the right to make any necessary as they become necessary Any changes made to our products may cause slight variations between the illustrations and explanations in this manual and the product you have purchased If you have any questions or concerns regarding the information provided in this manual please contact our Customer Service Department fo...

Page 3: ...ng the warnings cautions and information provided in this User Manual The symbols below are used throughout all Motion Concepts manuals to identify warnings cautions and important notes It is very important for you to read and understand them completely CAUTION Failure to heed the cautions in this Manual may result in damage to your Electronics and or your Power Positioning System WARNING Failure ...

Page 4: ...ch display to cycle through the available motor functions on the power positioning system and the Up Down button arrows to operate the selected motor function STATUS INDICATOR The screen display includes a color coded status bar and text at the top of the LCD screen display that will indicate the position status of the seating system with respect to safety limits and lockouts GREEN Status Indicato...

Page 5: ...PPS Owners Manual w TRD0512 UL Maxx UpFront Anterior Tilt Reach MPPS Owners Manual w TRD0527 Modular Power Standing System MPS Maxx Owners Manual 1 0 USING THE M290 REMOTE as an ATTENDANT CONTROL Scroll via arrows to select seat function ie g TILT RECLINE ELEVATE STAND LEG e g TILT FORWARD e g TILT BACKWARD 1 Use the LEFT RIGHT touch screen arrows 1 to scroll through the available power positionin...

Page 6: ...e available program options please contact your local Dealer or Service Provider for assistance 2 0 PROGRAM MODE for the M290 REMOTE 2 Note the M290 software version is also indicated briefly on the display screen when the remote attendant control is intially powered on PROGRAM OPTIONS1 DESCRIPTION Actuator1 Actuator motor functions may be individually programmed i e turned on and off Speed1 Sets ...

Page 7: ...rious injury or death Your power positioning system should never exceed the recommended safety limits These Safety Limits are controlled by a Tipsy Angle Sensor that is mounted in a fixed position on our Motion Concepts Power Positioning Systems The angle at which the limits are set is critical to the safe operation of the power positioning system refer to the recommended Safety Limts Angles provi...

Page 8: ...USA 700 Ensminger Rd Suite 112 Tonawanda NY USA 14150 716 447 0050 888 433 6818 Canada 84 Citation Dr Units 1 7 Concord Ontario Canada L4K 1C9 905 695 0134 866 748 7943 www motionconcepts com ...

Reviews: