System Control Panel
24
Logfile
5.3.3
-
Press the button „LOG“
-
The logfile will be displayed
User setup
5.4
In the user setup values such as limitations for the conductivity sensors, tank sensor or
system time setting can be adjusted.
-
To enter the user setup please enter the password
9999
and confirm entry with “OK”
Limits setup
5.4.1
-
Press the button „Limits”
-
“Set Q3-limit”:
That’s the maximum value of the
conductivity Q3 without showing any warning on the
screen
-
“Display Q”:
Switching the display from µS/cm to
MΩ*cm
-
“Press.Switch ON/OFF”:
The monitoring of the inlet
water pressure sensor can be switched on/off
-
“Interval ON/OFF”:
is to define how the system
works when the tank is already full
-
Interval ON: system recirculates every 3 hours
-
Interval OFF: system recirculates permanently