Echo/Vascular User Manual (15381-C)
Page 14 of 46
DCR-00172
Use Instructions
Powering the Product
The product can be powered by AC power from a wall outlet or by DC power via the
optional battery.
The product is “on” whenever it is plugged into AC power or when a
charged battery is installed. The product should not be positioned in a way that would
make it difficult to remove power by unplugging the AC power cord or removing the
battery.
To apply AC power to the product, attach
the power cord to the AC input located at
the rear of the product.
To apply DC power to the product, install a
battery at the foot end of the product.
Note
: The battery may be charged while it
is installed in the product by plugging the
product into AC power.
WARNING
•
To reduce the risk of electrical shock, grounding reliability can only be achieved when the
equipment is connected to an equivalent receptacle marked “hospital only” or “hospital
grade.
”
•
If damage has occurred to the power cord, immediately remove the cord from service.
Failure to do so could result in serious injury or death.
•
If damage has occurred to the battery, immediately remove the battery from service.
Failure to do so could result in serious injury or death.
•
The battery should be periodically inspected for damage. Replace the battery if necessary.