Echo/Vascular User Manual (15381-C)
Page 33 of 46
DCR-00172
Troubleshooting Guide
SYMPTOM
PROBABLE CAUSE
SUGGESTION
No Actuator Function
Actuator(s) Not Running.
•
Power cord not plugged all
the way into wall receptacle.
•
Push power cord securely
into receptacle.
•
Power outlet receptacle not
supplying 120 VAC power.
•
Check power availability or
plug unit into another
receptacle.
•
The power cord may be
separated from the control
box.
•
Securely press power cord
into control box.
•
Battery may be drained.
•
Replace or charge battery.
•
Actuator cord may be
unplugged.
•
Push actuator cords securely
into actuator receptacle.
•
Product was overloaded and
tripped internal fuse in
control box.
•
Replace control box.
•
Product motion is limited at
certain heights and angles.
•
Return the surface to a level
position to restore full range
of motion
•
The product will not move
when hand control buttons
are pressed.
•
The hand control buttons
may be locked. If the green
LED on the hand control is lit
then press the unlock button
for 3 seconds to unlock the
hand control functions.
Squeaking noises during
operation.
•
Actuator pins are not
sufficiently lubricated
•
Apply WD40 or similar
lubricant to actuator pins.
WARNING
•
Do not modify this equipment without authorization of the manufacturer.
•
Removal of secured covers may increase the risk of electrical shock. Refer servicing to
qualified personnel.
•
The battery should be periodically inspected for damage. Replace the battery if necessary.