Specifications
(MAN-0013-A)
Page 38 of 47
DCR-00422
Environmental Conditions
ATTRIBUTE
Range for Use
Range for Storage and
Transport
Ambient Temperature
+5° to 40° C
-10° to +40° C
Relative Humidity
20% to 80% @ 30°C
non-condensing
20% to 80% @ 30°C
non-condensing
Atmospheric Pressure
700 to 1060 hPa
700 to 1060 hPa
Upholstery
ATTRIBUTE
SPECIFICATION
Foam
California Technical Bulletin 117
Vinyl
California Technical Bulletin 117
Power Requirements
ATTRIBUTE
SPECIFICATION
Electrical, Product
100
–
240 VAC, 50/60 Hz, max 4.5 A
Battery Option
25.9V, 2.25Ah, 58.28Wh
Duty Cycle
10% max, 2 min. on / 18 min. off
Battery Duty Cycle
5% max, 1 min on / 19 min off
•
All electrical circuitry is isolated from chassis.
•
Grounding reliability can only be achieved when the equipment is connected to
an equivalent receptacle marked “Hospital Only” or “Hospital Grade
.
”
•
The power cord is to be used for mains disconnection.
•
Attached power supply cord set is not allowed to be diverted to other
equipment.