(MAN-0013-A)
Page 7 of 47
DCR-00422
Intended Use
This product is intended to be used in an environment where ultrasound and diagnostic
equipment is present, including hospitals, and outpatient facilities. The product is
intended to be used by healthcare professionals who possess the ability to operate the
product safely. The product’s movements are controlled both manually and electronically
via the
product’s
hand control.
The product is not intended for use in oxygen rich environments.
Safety Features
•
This product is equipped with multiple automated safety features to prevent danger
or damage during use. The entire system is electrically isolated to UL/IEC 60601-1 and
CAN/CSA C22.2 No. 60601-1:14 basic safety standards.
•
The actuator assemblies are current overload protected. If overloaded, the actuators
will stop and reset automatically.
•
The sealed hand control operates the actuators by directing small amounts of low
voltage D.C. current to the control box. All actuator drives are equipped with internal
limit switches which automatically prevent over-extension.