VOXI User’s Guide
15
5.4.2 Configure Pump Stall Threshold
This feature protects the pump from burning out if there is a
flow blockage that is causing excessive strain on the pump.
Click "Start Setting" to show the pump current draw.
Then block the inlet and observe the current draw again.
Enter the average value between the blocked and unblocked
pump, and click "Set and Exit" to write the updated threshold
to the VOXI. If this threshold is exceeded, the pump will stop
and an alarm signal sent.
5.4.3 Configure 4 mA Output Baseline
If sending the 4-20 mA signal to a remote controller, during
the field installation the VOXI analog output baseline should
be adjusted. Click the button under "4 mA" to set the output.
Click the increase and/or decrease buttons to adjust and send
to the VOXI immediately.
: Click +1; Long Click +10.
: Click +100; Long Click +1000.
: Click -1; Long Click -10.
: Click -100; Long Click -1000.
5.4.4 Wireless Registration
The VOXI can be connected (registered) to a wireless network for communication up to 2 miles
(3 km) away using an mLink modem. First, put the mLink in the "Register Device" mode
(
NOTE
:
this requires a separate mobile phone or tablet dedicated to the mLink
). Then
click the "Start Register" button on the App to register this VOXI to the mLink. Registration
should be completed immediately, and the results displayed in the pane to the right of the button.