09/ 2007
7.5 Resistance measuring ranges
(Switch setting:
Ω
)
Overload protection in the case of resistance measurements: 600 V
eff
Meas. range
Resolution
Meas. precision
Max. idling voltage
200 Ω
10 mΩ
± (0.3 % of the measured value + 30 Digit)
3,3 V
2k
Ω
100 mΩ
± (0.3 % of the measured value + 30 Digit)
3,3 V
20 kΩ
1 Ω
± (0.3 % of the measured value + 30 Digit)
3,3 V
200 kΩ
10 Ω
± (0.3 % of the measured value + 30 Digit)
3,3 V
2 MΩ
100 Ω
± (0.3 % of the measured value + 30 Digit)
3,3 V
20 MΩ
1 kΩ
± (5.0 % of the measured value + 50 Digit)
3,3 V
200 MΩ
1 MΩ
± (5.0 % of the measured value + 20 Digit)
3,3 V
2G
Ω *
2
0,1 GΩ
± (5 % of the measured value + 8 Digit)
3,3 V
*
2
The 2 G
Ω
measuring range has to be selected manually by means of the
key
L
(see item 5.1.1.4).
7.6 Resistance measurement range with reduced measuring voltage: 0.6 V,
(Switch setting: LV
Ω
, indication in display “LV”)
Meas. range
Resolution
Meas. precision
Max. idling voltage
2k
Ω
0,1 Ω
± (0.6 % of the measured value + 30 Digit)
0,6 V
20 kΩ
1 Ω
± (0.6 % of the measured value + 30 Digit)
0,6 V
200 kΩ
10 Ω
± (0.6 % of the measured value + 30 Digit)
0,6 V
2 MΩ
100 Ω
± (0.6 % of the measured value + 30 Digit)
0,6 V
20 MΩ
1 kΩ
± (7.0 % of the measured value + 50 Digit)
0,6 V
200 MΩ
1 MΩ
± (7.0 % of the measured value + 20 Digit)
0,6 V
7.7 Diode and continuity testing
(Switch setting:
, )
Overload protection: 600 V
eff
The built-in buzzer sounds in the case of a resistance R less than 50
Ω.
Meas. range
Resolution
Max. meas. current
Max. idling voltage
1 mV
1,1 mA
3,3
V
7.8 Capacity ranges
(Switch setting: )
Largest display value: 4.000 Points
Conditions: Discharge capacitors and apply according to the specified polarity.
Meas. range
Resolution
Meas. precision
Overload protection
4nF
1 pF
± (1.5 % of the measured value + 10 Digit)
600 V
eff
40 nF
10 pF
± (1.5 % of the measured value + 10 Digit)
600 V
eff
400 nF
100 pF
± (0.9 % of the measured value + 5 Digit)
600 V
eff
4 µF
1 nF
± (0.9 % of the measured value + 5 Digit)
600 V
eff
40 µF
10 nF
± (1.2 % of the measured value + 5 Digit)
600 V
eff
400 µF
100 nF
± (1.2 % of the measured value + 5 Digit)
600 V
eff
4mF
1 µF
± (1.5 % of the measured value + 5 Digit)
600 V
eff
40 mF
10 µF
± (1.5 % of the measured value + 5 Digit)
600 V
eff
7.9 Frequency ranges
(Switch setting: Hz)
Overload protection in the case of frequency measurements: 600 V
eff
Minimum input frequency: 5 Hz
Meas. range Resolution
Meas. precision for 5 V
eff
max.
Min. sensitivity
20 Hz
0,001 Hz
± (0.01 % of the measured value + 50 Digit)
0,5 V
SS
200 Hz
0,01 Hz
± (0.01 % of the measured value + 10 Digit)
0,5 V
SS
2kHz
0,1 Hz
± (0.01 % of the measured value + 10 Digit)
0,5 V
SS
20 kHz
1 Hz
± (0.01 % of the measured value + 10 Digit)
0,5 V
SS
200 kHz
10 Hz
± (0.01 % of the measured value + 10 Digit)
0,5 V
SS
1 MHz
100 Hz
± (0.01 % of the measured value + 10 Digit)
0,5 V
SS