Recirculating cooler
(
Chiller
)
6
Location Principle
1
Firm ground or experiment table in a dust-free and cool environment with
good ventilation
2
Far from droplets or steam
3
Far from the location that influx, viscous flow, leak of fuel gas may happen
4
Far from high frequency devices such as high frequency welding machine
5
Far from the location where acidic solutions used frequently
6
Far from the location where special sprayer (sulfide contained) used
frequently
7
Ambient temperature 5 ° to 30 ° C, relative humidity less than 80%
(non-condensing)
As for convenience and safety of maintenance and installation, keep certain
distance between the indoor unit and obstacles.
Ensure the measurements during indoor installation:
1
)
The distance between the bath top and obstacles shall be at least 300mm;
2
)
The distance between two bath sides and obstacles shall be at least 500mm;
3
)
The distance between bath back and obstacles shall be at least 500mm;
4
)
The distance between bath front and obstacles shall be at least 1000mm;
Such as following picture:
(unit: mm)