Recirculating cooler
(
Chiller
)
28
Appendix I Parameter Database
Temperature set menu: no password, press
【
menu
】
key can enter directly;
Item
Parameter Signification
Option or Range
Initial data
Target
temperature
T_00
Setting the target
temperature
(Start/Stop type
,
PID)
C_SL ~C_SH
20
Temperature
deviation
T_01
Setting
control
precision
(Start/Stop type)
0.01 ~ 9.99
5
Tracking
temperature
T_11
Setting deviation
temperature
(tracking mode)
-99.9-99.9
-2
Temperature correction database: password 0011, long
【
menu
】
key enter the
password interface input 0011 to enter;
Item
Parameter Meaning
Option or Range
Initial data
Offset correction R_TB
Temperature
correction offset
-199.9~999.9
0
Slope correction R_TK
Temperature
correction slope
0.001~9.999
1
Temperature
display set
R_CF
Temperature display
options
F -
℉
C
C -
℃
Digital filtering
R_FT
Digital filtering time 0.0~120.0S
S
End
_END
Parameter set end
signal
----
----