Recirculating cooler
(
Chiller
)
33
Appendix II Warning Information
Item
Alarm lamp and
buzzer
Auxiliary
display
Main display
Solution
High temperature
alarm 1
2S flash
E_H1
Current
temperature
No,
quit
when
the
temperature lower than
E_H1, back to original work
mode
High temperature
alarm 2
1S flash
E_H2
Current
temperature
No,
quit
when
the
temperature lower than
E_H2, back to original work
mode
Low temperature
alarm 1
2S flash
E_L1
Current
temperature
No,
quit
when
the
temperature higher than
E_L1, back to original work
mode
Low temperature
alarm 2
1S flash
E_L2
Current
temperature
No,
quit
when
the
temperature higher than
E_L2, back to original work
mode
Water level alarm 2S flash
E_LS
Current
temperature
Stop heating/cooling, quit
at normal water level and
back to original work mode
Water flow alarm 1S flash
E_LF
Current
temperature
Stop heating/cooling, quit
at normal water level and
back to original work mode
Sensor 1 break 1S flash
E_S1
999
Stop heating/cooling, quit
at normal water level and
back to original work mode
Sensor 1 short
circuit
1S flash
E_S1
-999
Stop heating/cooling, quit
at normal water level and
back to original work mode
Sensor 2 break 1S flash
E_S2
999
Stop heating/cooling, quit
at normal water level and
back to original work mode
Sensor 2 short
circuit
1S flash
E_S2
-999
Stop heating/cooling, quit
at normal water level and
back to