6.3 Configuration of measurement mode
Configuration of stability sense and data storage method
While pressing [esc] key, press [ok/meas.] key to show Function screen.
Select <Measure Mode> with [
+
], [
−
] key.
Press [ok/meas.].
Custom
Field
Labo
Sample Name
: Single alphabet (space, a to z) can be used for identification.
Stability
: Setting automatic stability sense mode (Auto, Manu)
[Auto]
Checks stability of density of sample automatically. If stabilized, the measurement
data is automatically retained.
[Manu] After
measurement,
press
[ok/meas.] key to hold desired measurement data
Mode
: Setting automatic stability sense mode (Custom, Labo, Field)
[Custom]
Selection of mode for automatic saving retained measurement results or output
externally.
You can select Memory in or Memory out as follows:
Memory in : Setting storage method of measurement results
[Auto] When measurement results are retained, the value is automatically saved.
[Manu] After measurement results are retained, press [ok/meas.] key to save desired
measurement data
Memory out : Setting output method of measurement results
[Auto] The measurement results are retained and the value is automatically printed out
simultaneously.
[Manu] Press [mem.out/all] key to output the measurement data.
[Labo]
The retained data is manually saved with [ok/meas.] key and automatically output
externally.
Memory in to set to “Manu” and Memory out to “Auto” automatically.
[Field]
The retained data is manually saved with [ok/meas.] key.
Memory in to set to “Manu” and Memory out to “Auto” automatically.