30
9.5.2 Data output to Printer
In order to output data to an external printer, it is necessary to connect the optional infrared RS converter
(98-029-0007) and polar converter (98-404-0007). The optional stand (98-029-0006) will help you
easily set up infrared receiving unit of DA-130N. An example of setting the stand and an IDP-100
printer is described here in this manual. Install the main unit and infrared RS converter onto the stand as
shown, and connect the polar converter (98-404-0007) to IDP-100 Printer, and then, connect the infrared
RS converter.
Stand
(98-029-0006)
Infrared RS converter
(98-029-0007)
Polar converter
(98-404-0007)
Printer
(IDP-100)
Select “PRN”(printer) on <Interface> of Function screen to make communication settings.
ok
•
The optional infrared RS converter (98-029-0007) works only at 9600 baud.
Therefore, “Baud Rate” must be set at “9600.”
•
Communication with IDP-100 printer also works only at 9600 baud.
Printout Sample of Measurement Results:
The marked data will
appear with "*" symbol
as a suffix to sample
number.
Sample No. 0001
Date : 2003/10/28 14:03
0.9972 g/cm3
25.1 C
ID : SAMPLE-001
Sample No. 0002
Date : 2003/10/28 14:08
0.9965 g/cm3
25.3 C
ID : SAMPLE-001
Sample No. a0003
Date : 2003/10/28 14:13
0.9977 g/cm3
24.8 C
ID : SAMPLE-001
Sample No. b0004
Date : 2003/10/28 14:18
0.9980 g/cm3
24.5 C
ID : SAMPLE-001
Sample No. b0005*
Date : 2003/10/28 14:23
0.9910 g/cm3
24.5 C
ID : SAMPLE-001
Sample No. b0006
Date : 2003/10/28 14:28
0.9956 g/cm3
24.9 C
ID : SAMPLE-001