41
15.2 Exhibit no repeatability on measured value or show deviation on measured value
START
Is sample uniform?
Stir sample to make it uniform
(Sample with separated lipuid phase
results in deviating measurements)
Yes
No
Has calibration been
performed properly?
Does air leak form the joints?
Are air bubbles
in sample?
Has measureing cell stain?
Take another sample
by proper sampling
Check 'Calib. Mode' and
calibrate again
Check and see
the joints for leakage
Rinse measuring cell well
Yes
Yes
Yes
Yes
No
No
No
No
Contact our sales
representative
Is everything O.K.?
END
Yes
No