background image

1

  Safety notes 

·

  Transport the equipment with care!

·

  Careful transportation of the equipment requested !

· 

The unit may NEVER be overturned nor put upside down !

· 

Equipment and its internal parts can be damaged :

- by dropping
- by shock.

·

  Do not start up the equipment if Equipment should only be operated by

properly instructed personnel ! 

- it is damaged or leaking
- the supply cable is damaged.

·

Switch off the equipment and pull out the mains plug for:

- servicing or repair

- before moving the equipment!

·

Have  the  equipment  serviced  or  repaired  only  by  properly  qualified

personnel ! 

· 

At low temperatures, do not move or bend    the insulated flex line tube !

·

  Be careful not to bend the insulated flex line appreciably after the unit has

chilled down the probe and line. 

·

  The flexible line to the cold probe should have no sharp bends nor signs

of broken tubing. 

2

 

Unpacking 

Goods are packed carefully, largely preventing transport damage. 
After  unpacking  ,  check  first  the  equipment  and  accessories  for  possible 

transport damage. If unexpectedly any damage can be noticed on the equipment it is 
essential to notify the forwarding agent or the postal authorities so that an inspection 
can take place. 

Do not start operations with damaged equipments! 
The unit may NEVER be overturned nor put upside down ! 

3

  Equipment description

 

Refrigeration unit 

The refrigeration unit consists essentially of a fully sealed compressor. The Cryocoil 
or Cryoprobe is cooled down to a very low temperatures,through a flexible heat 
exchanging tube located on the top of this equipment. Heat of condensation and 
motor heat are dissipated on air-cooled units through a fan-cooled ribbed condenser. 

Summary of Contents for FIGR-100

Page 1: ...ctions and safety notes If you have any questions please phone us Follow the instructions on setting up operation etc This is the only way to exclude incorrect operation of the equipment and to ensure...

Page 2: ...n the probe and line The flexible line to the cold probe should have no sharp bends nor signs of broken tubing 2 Unpacking Goods are packed carefully largely preventing transport damage After unpackin...

Page 3: ...r d throttling and evaporating this latter condensate in a coil to produce cooling at very low temperatures e passing the refrigerant leaving the cooling coil back through the heat exchanger to the co...

Page 4: ...sures Control unit The control unit with indication and operating functions The equipment is provided with a LED display to indicate the measurements and settings as well as the operational status The...

Page 5: ...off Remove the liquid nitrogen cold trap from the vacuum system and insert the cold probe in its place Make certain there are no sharp bends or strains on the flexible insulated line leading to the co...

Page 6: ...ess Field Parameter Table Pressing can go to the next parameter pressing or can modify a parameter The equipment will escape auomatically from the parameter table if no key is pressed within 30 second...

Page 7: ...ial 5 Maintenance 5 1 Maintenance and Periodic Inspections The only component requiring maintenance is the air cooled condenser It should be kept clean enough for free air flow through it Accumulated...

Page 8: ...servicing it is advisable to contact our Technical Service department If the equipment has to be returned to the factory please ensure that it is carefully and properly packed MRC accepts no responsib...

Reviews: