background image

value first, and then start auto-tuning. Generally, for application with heating time of 10%

80% 

can obtain satisfying auto-tuning result. If the heating time is not in the range of 10

80%, the 

auto-

tuning will run for a long time and can’t get satisfying result. Raising setpoint for heating 

time less than 10% or decreasing setpoint for heating time greater than 80% can improve 

auto-tuning result. 

Note 2: 

Don’t modify the setpoint during auto-tuning. It may affect the precision of auto-tuning. 

Note 3: 

The temperature fluctuation caused by interference may affect the precision of 

auto-tuning. Besides checking the wiring, increasing the parameter FILt (input digital filter) can 

also reduce interference. 

Note 4: 

After auto-tuning finishes, generally, the temperature rising has a little over-shoot (about 

1~3

). It is because considering that the transducer is often placed between the heater and 

heated material, and a little over-shoot can shorten the time of rising to setpoint and save power. 

(Generally, the temperature of transducer can be raised quick than that of heated material) 

5 Maintenance 

5.1 Maintenance and Periodic Inspections 

The only component requiring maintenance is the air-cooled condenser. It should be 
kept clean enough for free air flow through it. Accumulated dirt or lint may be 
removed by brushing, 
vacuum cleaning or with a compressed air blow-gun. Inspect the condenser face 
about weekly for the first month then monthly the rafter to determine the cleaning 
frequency which will be required. 
The compressor and condenser fan motors are lifetime lubricated and do not require 
regular oiling. If practical, the temperature of the compressor should be checked 
every several months by feeling it. It should be possible to hold a hand on top of it for 
a few seconds. 
Also there should be no frost accumulation on the suction valve on the side of the 
compressor.

 

5.2

   

Cleaning

 

Before cleaning the equipment, pull out the mains plug!

 

The unit can be cleaned with water with the addition of a few drops of a detergent 
(washing-up liquid), using a moist cloth. 
Water must not enter the control unit 
The user is responsible for carrying out an appropriate detoxification of any 
dangerous material which has been spilled on or in the unit. This applies in particular 
when the unit is passed on to someone else for operation, repair, storage etc.. 

Summary of Contents for FIGR-100

Page 1: ...ctions and safety notes If you have any questions please phone us Follow the instructions on setting up operation etc This is the only way to exclude incorrect operation of the equipment and to ensure...

Page 2: ...n the probe and line The flexible line to the cold probe should have no sharp bends nor signs of broken tubing 2 Unpacking Goods are packed carefully largely preventing transport damage After unpackin...

Page 3: ...r d throttling and evaporating this latter condensate in a coil to produce cooling at very low temperatures e passing the refrigerant leaving the cooling coil back through the heat exchanger to the co...

Page 4: ...sures Control unit The control unit with indication and operating functions The equipment is provided with a LED display to indicate the measurements and settings as well as the operational status The...

Page 5: ...off Remove the liquid nitrogen cold trap from the vacuum system and insert the cold probe in its place Make certain there are no sharp bends or strains on the flexible insulated line leading to the co...

Page 6: ...ess Field Parameter Table Pressing can go to the next parameter pressing or can modify a parameter The equipment will escape auomatically from the parameter table if no key is pressed within 30 second...

Page 7: ...ial 5 Maintenance 5 1 Maintenance and Periodic Inspections The only component requiring maintenance is the air cooled condenser It should be kept clean enough for free air flow through it Accumulated...

Page 8: ...servicing it is advisable to contact our Technical Service department If the equipment has to be returned to the factory please ensure that it is carefully and properly packed MRC accepts no responsib...