27
Image Setup Sub-menu
Alarm Switch
Press
or
turns on or off the alarm function. If the Alarm Switch is
on, when target subject within area box has surface temperature over
Alarm Temp, device will alarm with buzzer sound and mark the high
temperature location as Alarm Color if Alarm Color is set.
Alarm Temp
When Alarm Switch is off, this option is disabled. When Alarm Switch is on,
Alarm Temp is the desired alarm temperature. Press
or
once to
change Alarm Temp 0.1
. Long press
or
will change Alarm
Temp in 1
interval. Default setting is 37.0
.
Alarm Color
Alarm Color can select as: Auto, Black, White, Red, Orange, Yellow,
Green, Blue, Gray, and Purple. Default setting is Auto, which is no color
alarm.
Isotherm Color
When this option is Auto, there is no Isothermal display. Otherwise, target
subject Infrared image will change to selected color when temperature is
within the range of (Isothermal Temp ± Isothermal Width/2). Default setting
is Auto.
Isotherm Temp