USER MANUAL SWG
Page 10 of 129
are included, each operating with a different optical path length and
stabilized by referring to a reference detector. The IR source is a
highly efficient and stable IR emitter, pulsed at a frequency of several
Hertz. By design NDIR technique offers good stability and selectivity
together with long lifetime of sensor (only limited by corrosion or
dust, which can be prevented or removed by regular servicing the
instrument).
•
The instrument is optionally equipped with several electro-chemical
sensors ECS to detect gas components like oxygen
O
2
. Those sensors
offer a reliable and effective way to detect the target gases. They are
typically of limited lifetime (several years) but may be easily replaced
once the end-of-life is reached.
•
In regular time intervals the instrument automatically switches to
purge the sensors with fresh (ambient) air for re-adjust the zero point.
Picture above: Sample gas flow diagram (Picture is an example. It only shows the
principle).
Summary of Contents for SWG 100 SYNGAS
Page 1: ...SWG 100 SYNGAS USER MANUAL ...
Page 12: ...USER MANUAL SWG Page 12 of 129 11 H3PO4 inlet G1 8 inner thread ...
Page 24: ...USER MANUAL SWG Page 24 of 129 ...
Page 32: ...USER MANUAL SWG ...
Page 63: ...USER MANUAL SWG Page 63 of 129 Detail A Alarm Relais ...
Page 89: ...USER MANUAL SWG Page 89 of 129 8 3 Description of the analyser 1 1 1 1 2 ...
Page 99: ...USER MANUAL SWG Page 99 of 129 ...
Page 110: ...USER MANUAL SWG Page 110 of 129 12 Appendix ...
Page 114: ...USER MANUAL SWG Page 114 of 129 ...
Page 116: ...USER MANUAL SWG Page 116 of 129 Positions of the components 3 4 ...
Page 127: ...USER MANUAL SWG Page 127 of 129 ...
Page 129: ...USER MANUAL SWG Page 129 of 129 ...