USER MANUAL SWG
Page 107 of 129
•
Open adjustment gas cylinder: Open the adjustment gas cylinder.
The actual value changes. After 1 until 2 minutes the values should
no longer change. If the value is adjustable, the set value is shown
in quotation marks.
•
Set a new adjustment factor: Finally, pressing the OK key sets the
new adjustment factor.
•
During the whole adjustment procedure, only the lower setpoint
value is changed. The upper set point remains unchanged. This is
only adjusted if the accuracy in the upper range is to be improved.
open
P=500 mbar