The pressure reducer unit must be installed at the sample gas inlets. It is a part of
the safety concept of the analyzer, which must be observed.
Correct installation of the pressure reducer unit:
•
Install a copper seat to the pressure reducer unit.
•
Use PTFE strip for the outer thread of the pressure reducer unit.
•
Screw the pressure reducer unit into the sample gas inlet. Use a
19 open end spanner.
Picture: Flow restrictor orifice with intern filter.
Reference:
1. Sample gas inlet.
2. Calibration gas inlet max.4
1
Sinter filter
O-ring
Nozzle
Male connector G1/8
2