Detail B:
Reference:
4. Condensate outlet with G1/8 inner thread and connector for DN4/6 mm hoses
5. Sample gas outlet with G1/4 inner thread and connector for DN12/6 mm hoses.
Installation calibration gas inlet (see detail A-3)
•
WARNING
Overpressure
Do not set the pressure reducer over 500 mbar
•
WARNING
As part of the safety concept, a reducing unit must be
installed on each Autocal input.
Reference:
1. Reducing unit
2. Copper seal.
The calibration gas inputs are normally located on the right side, the upper gas
connection plate. Optionally, 4 can be installed. Each calibration gas port is
intended for one calibration gas cylinder and has a screw connection for DN4/6 mm
hoses.
The pressure regulator must not be set to more than 500 mbar.
Installation zero gas inlet (see detail A-2)
The zero-gas connection is located on the left side of the upper gas connection
plate and is equipped with a screw connection for DN4/6 mm hoses. The zero-gas
connection is used to draw ambient air into the analyser at defined periods and
thus to take the zero point for the internal measuring sensors.
The zero gas should be taken from a place where no exhaust gases are emitted, so
that the analyser can take the optimum zero point.
4
5
1
2
Summary of Contents for SWG 100 SYNGAS
Page 1: ...SWG 100 SYNGAS USER MANUAL ...
Page 12: ...USER MANUAL SWG Page 12 of 129 11 H3PO4 inlet G1 8 inner thread ...
Page 24: ...USER MANUAL SWG Page 24 of 129 ...
Page 32: ...USER MANUAL SWG ...
Page 63: ...USER MANUAL SWG Page 63 of 129 Detail A Alarm Relais ...
Page 89: ...USER MANUAL SWG Page 89 of 129 8 3 Description of the analyser 1 1 1 1 2 ...
Page 99: ...USER MANUAL SWG Page 99 of 129 ...
Page 110: ...USER MANUAL SWG Page 110 of 129 12 Appendix ...
Page 114: ...USER MANUAL SWG Page 114 of 129 ...
Page 116: ...USER MANUAL SWG Page 116 of 129 Positions of the components 3 4 ...
Page 127: ...USER MANUAL SWG Page 127 of 129 ...
Page 129: ...USER MANUAL SWG Page 129 of 129 ...