4.5.
Analog inputs (4-20 mA)
Hardware-side:
The analog inputs are on the top of the IO module. They are marked with a blue
label. Through the help of the analog inputs, the IO module can read all
common 4-20 mA transducer in the analyzer directly. The IO module has a
separate 12 V power source, for the supply of the transducer. At the sketch
below, the inputs and the power supply, for the transducer are marked.
Information:
Internal Load:
50 Ohm
Power supply for transducer:
12VDC / 200 mA
The sketches below show how the common transducer can connect to the IO
module.
V12+
AI1
+
AI2
+
AI1
-
AI2
-
AI3
+
AI4
+
AI3
-
AI4
-
Power supply for transducer
Analog inputs
GND