USER MANUAL SWG
Page 62 of 129
5.5.
Configuration of the alarm relays
contact.
The following errors will turn the relay from NC to NO.
1.
Internal RS485 bus communication failure
2.
Analyser is in bootloader phase
3.
Gas leakage inside analyser cabinet (CH4 > 20% to 50% LEL)
4.
Condensate alarm (contacts resistance < 35k )
5.
Low fan rotation (speed rotation < 900U/min)
6.
Sample flow alarm (sample flow < 20 l/hr)
7.
Gas cooler high alarm (temperature > +10°C)
8.
Gas cooler low alarm (temperature < +2°C)
9.
Cabinet high temperature (> +55°C)
10.
Cabinet low temperature (< +5°C)
Errors
1 to 5
alarm will force a measurement stop (all analog outputs are on hold
or at 2mA, depending on configuration).
Errors
6 to 10
will be displayed as warning message only; analog outputs of
active sampling point are live all others are on hold.
NOTE
Analyser system alarm relay is a potential free contact, with
max 300 VAC / 10 A or 150 VAC / 12 A.
A
Summary of Contents for SWG 100 SYNGAS
Page 1: ...SWG 100 SYNGAS USER MANUAL ...
Page 12: ...USER MANUAL SWG Page 12 of 129 11 H3PO4 inlet G1 8 inner thread ...
Page 24: ...USER MANUAL SWG Page 24 of 129 ...
Page 32: ...USER MANUAL SWG ...
Page 63: ...USER MANUAL SWG Page 63 of 129 Detail A Alarm Relais ...
Page 89: ...USER MANUAL SWG Page 89 of 129 8 3 Description of the analyser 1 1 1 1 2 ...
Page 99: ...USER MANUAL SWG Page 99 of 129 ...
Page 110: ...USER MANUAL SWG Page 110 of 129 12 Appendix ...
Page 114: ...USER MANUAL SWG Page 114 of 129 ...
Page 116: ...USER MANUAL SWG Page 116 of 129 Positions of the components 3 4 ...
Page 127: ...USER MANUAL SWG Page 127 of 129 ...
Page 129: ...USER MANUAL SWG Page 129 of 129 ...