Input
Remote Alarm Reset
Normally open digital input
Power
Power consumption
2 watts (4 watts for 5100-88-IT)
Input voltage
24 VDC nominal: 10-30 VDC
Operating Range
Ambient Temp (°F) Ambient Temp (°C) Relative Humidity
5100-03-IT O
2
-4 to 122 °F
-20 to 50 °C
15 – 90%
5100-04-IT CO
-4 to 122 °F
-20 to 50 °C
15 – 90%
5100-05-IT H
2
S
(1)
- 40 to 140 °F
-40 to 60 °C
15 – 90%
5100-06-IT Cl
2
-4 to 122 °F
-20 to 50 °C
15 – 90%
5100-08-IT ClO
2
-4 to 122 °F
-20 to 50 °C
15 – 90%
5100-10-IT SO
2
-4 to 122 °F
-20 to 50 °C
15 – 90%
5100-12-IT NO
2
-4 to 122 °F
-20 to 50 °C
15 – 90%
5100-25-IT NH
3
+14 to 113 °F
-10 to 45 °C
20 – 95%
5100-26-IT HF
+14 to 113 °F
-10 to 45 °C
20 – 95%
5100-88-IT CO
2
- 40 to 140 °F
-40 to 60 °C
15 – 90%
(1) Performance tested by FM Approvals to -40°C/°F
Enclosure
Material - Aluminum
Polyester powder-coated, sand-cast, copper-free aluminum
Material - Stainless Steel
316 Stainless Steel
Dimensions (HxWxD) (differs
for 5100-25-IT & 5100-26-IT)
(A1 & A2) 7.9 x 5.7 x 3.9 in. (20.1 x 14.5 x 9.9 cm)
(S1 & S2) 7.6 x 4.4 x 4.9 in. (19.3 x 11.2 x 12.4 cm)
Weight
(A1 & A2) 3.4 lb. (1.5 Kg)
(S1 & S2) 7.2 lb. (3.3 Kg)
Electrical Approvals (5100-05-IT Only)
Factory Mutual (FM)
Class 1, Div. I, Groups B, C, D
FM Standards 3600, 3615, 3810
ATEX (5100-XX-IT)
II 2 G Ex d IIB+H2 T6 Gb Ta = -40°C to +50°C
SMC 5100-XX-IT Toxic Gas Detector Module
41
US
8 Specifications
Summary of Contents for 5100-03-IT
Page 5: ...15 5100 88 IT Dimensions 67 SMC 5100 XX IT Toxic Gas Detector Module 5 US...
Page 20: ...mA Circuit Types 20 SMC 5100 XX IT Toxic Gas Detector Module 4 Installation US...
Page 50: ...12 Modbus Memory Map 50 SMC 5100 XX IT Toxic Gas Detector Module 12 Modbus Memory Map US...
Page 51: ...13 HART SMC 5100 XX IT Toxic Gas Detector Module 51 US 13 HART...
Page 52: ...52 SMC 5100 XX IT Toxic Gas Detector Module 13 HART US...
Page 68: ......