ALTAIR
®
4XR
16
Description
US
Minimum Readings (MIN page)
This page shows the lowest level of oxygen recorded by the device since turn-on or since the MIN
reading was reset.
The minimum icon (
chapter 2.3 "On-Screen Indicators") appears on the display.
To reset the MIN reading:
(1)
Access the MIN page.
(2)
Press the
button.
Short Term Exposure Limits (STEL page)
WARNING!
If the STEL alarm activates, leave the contaminated area immediately; the ambient gas concentration
has reached the preset STEL alarm level. Failure to follow this warning will cause over-exposure to
toxic gases and persons relying on this product for their safety could sustain serious personal injury or
death.
The STEL icon (
chapter 2.3 "On-Screen Indicators") appears on the display to show the average
exposure over a 15 minute period.
When the amount of gas detected by the device is greater than the STEL limit:
•
Alarm sounds
•
Alarm LEDs flash
•
STEL icon flashes.
To reset the STEL:
(1)
Access the STEL page.
(2)
Press the
button.
The STEL alarm is calculated over a 15 minute exposure.
STEL calculation examples:
Assume the device has been running for at least 15 minutes:
15 minute exposure of 35 ppm:
10 minute exposure of 35 ppm and 5 minutes exposure of 5 ppm:
This page can be de-activated through MSA GALAXY GX2 Test Stand or MSA Link.
(15 minutes x 35 ppm)
= 35 ppm
15 minutes
(10 minutes x 35 ppm) + (5 minutes x 5 ppm)
= 25 ppm
15 minutes
Summary of Contents for ALTAIR 4XR
Page 1: ...MSAsafety com Operating Manual ALTAIR 4XR Multigas Detector Order No 10175896 00 ...
Page 52: ...ALTAIR 4XR 52 Appendix US 7 Appendix 7 1 Start Up Sequence Power on ...
Page 53: ...ALTAIR 4XR 53 Appendix US 7 2 Fresh Air Setup FAS ...
Page 55: ...ALTAIR 4XR 55 Appendix US ...
Page 56: ...ALTAIR 4XR 56 Appendix US 7 4 Bump Test ...
Page 57: ...ALTAIR 4XR 57 Appendix US 7 5 Options Setup ...
Page 58: ...ALTAIR 4XR 58 Appendix US 7 6 Sensor Setup ...
Page 59: ...ALTAIR 4XR 59 Appendix US 7 7 Calibrations ...
Page 60: ...ALTAIR 4XR 60 Appendix US ...
Page 61: ...ALTAIR 4XR 61 Appendix US 7 8 Alarm Setup ...
Page 62: ...ALTAIR 4XR 62 Appendix US 7 9 Time and Date Setup ...
Page 63: ...sales norrscope com ALTAIR 4XR 63 Appendix US 7 10 Bluetooth Setup ...