Operation
US
The
symbol will be displayed in the Normal Operation mode and the Bump LED will flash green for
24 hours after a successful Bump Test.
3.9
Bump LED
The device is equipped with a green Bump LED. The green LED flashes every 15 seconds under the
following conditions:
•
the Bump LED feature is enabled
•
after a successful Bump Test (for 24 hours)
•
device is in Normal Operation mode
•
device is not in Low Battery warning or alarm
3.10
Calibration
The device can be calibrated either manually using this procedure or automatically using the GALAXY
GX2 Test Stand. Refer to 7.7 "Calibrations".
Calibration must be performed using a flow regulator with a flow rate set to 0.25 liters per minute.
If a battery charging cycle is interrupted before it is completed (4 hours for a fully discharged battery),
allow the device’s internal temperature to stabilize for 30 minutes before performing a Calibration.
NOTE:
The GALAXY GX2 Test Stand is not a CSA certified calibration method.
Fresh Air Setup and Zero Calibration
(1)
Press and hold the
button in Normal Operation mode for three seconds.
(2)
If calibration lockout option is selected, enter password.
ZERO CAL?
screen is then displayed.
Under normal circumstances MSA recommends calibration at least every six months,
however, many European countries have their own guidelines. Please check your
national legislation.
To skip the Zero procedure and move directly to the Calibration Span procedure, push
the
button. If no button is pushed for 30 seconds, the device prompts user to perform
a Span calibration before returning to the Normal Operation mode.
Summary of Contents for ALTAIR 4XR
Page 1: ...MSAsafety com Operating Manual ALTAIR 4XR Multigas Detector Order No 10175896 00 ...
Page 52: ...ALTAIR 4XR 52 Appendix US 7 Appendix 7 1 Start Up Sequence Power on ...
Page 53: ...ALTAIR 4XR 53 Appendix US 7 2 Fresh Air Setup FAS ...
Page 55: ...ALTAIR 4XR 55 Appendix US ...
Page 56: ...ALTAIR 4XR 56 Appendix US 7 4 Bump Test ...
Page 57: ...ALTAIR 4XR 57 Appendix US 7 5 Options Setup ...
Page 58: ...ALTAIR 4XR 58 Appendix US 7 6 Sensor Setup ...
Page 59: ...ALTAIR 4XR 59 Appendix US 7 7 Calibrations ...
Page 60: ...ALTAIR 4XR 60 Appendix US ...
Page 61: ...ALTAIR 4XR 61 Appendix US 7 8 Alarm Setup ...
Page 62: ...ALTAIR 4XR 62 Appendix US 7 9 Time and Date Setup ...
Page 63: ...sales norrscope com ALTAIR 4XR 63 Appendix US 7 10 Bluetooth Setup ...