ALTAIR
®
4XR
36
Operation
US
•
If calibration lockout option is NOT selected:
ZERO
screen displays.
(3)
With the device exposed to fresh air, press the
q
button to confirm the
ZERO CAL?
screen. A
sensor refresh and Zero Calibration will occur.
Alternatively press the
button to execute a Fresh Air setup (FAS). See chapter 3.2
"Turning ON and Fresh Air Setup" for more details.
After Zero calibration completes, the device momen-
tarily displays “
ZERO PASS
” or “
ZERO ERR
” along
with the flag of any sensor that failed.
During zero calibration the O
2
sensor is also span calibrated to 20.8% O
2
fresh air,
adjusting the calibration curve as needed. During span calibration, the O
2
sensor's accu-
racy is checked against a known oxygen gas concentration without adjusting the calibra-
tion curve.