Maintenance
US
(8)
Calibrate the device after the sensors have stabilized.
WARNING!
Calibration is required after a sensor is installed; otherwise, the device will not perform as expected
and persons relying on this product for their safety could sustain serious personal injury or death.
4.3
Cleaning
•
Routine Cleaning:
Clean the exterior of the device regularly using only a damp cloth. Do not use
cleaning agents, as many contain silicone, which will damage the combustible sensor.
•
Dust and Dirt Exposure:
Use a dry, soft bristled brush to remove any dust or dirt that has accu-
mulated on the apparatus, especially at the sensor openings. If there is a buildup of dust or dirt
particles remaining in the sensor area after brushing, use a vacuum to remove remaining particles,
but maintain at least a 1/2 inch (1.3 cm) distance from the gas detector.
•
Water Exposure:
If the apparatus is exposed to water, turn the device sensor side down and
gently shake water off the sensor area. Any remaining water can be removed with a clean dry
cloth.
4.4
Storage
When not in use, store the device in a safe, dry place between 18 °C (64 °F) and 30 °C (86 °F). After
storage, always recheck device calibration before use.
4.5
Scope of Delivery
Pack the device in its original shipping container with suitable padding. If the original container is
unavailable, an equivalent container may be substituted.
Allow sensors to stabilize at room temperature for at least 30 minutes before calibration
(
chapter 3.10 "Calibration").
Summary of Contents for ALTAIR 4XR
Page 1: ...MSAsafety com Operating Manual ALTAIR 4XR Multigas Detector Order No 10175896 00 ...
Page 52: ...ALTAIR 4XR 52 Appendix US 7 Appendix 7 1 Start Up Sequence Power on ...
Page 53: ...ALTAIR 4XR 53 Appendix US 7 2 Fresh Air Setup FAS ...
Page 55: ...ALTAIR 4XR 55 Appendix US ...
Page 56: ...ALTAIR 4XR 56 Appendix US 7 4 Bump Test ...
Page 57: ...ALTAIR 4XR 57 Appendix US 7 5 Options Setup ...
Page 58: ...ALTAIR 4XR 58 Appendix US 7 6 Sensor Setup ...
Page 59: ...ALTAIR 4XR 59 Appendix US 7 7 Calibrations ...
Page 60: ...ALTAIR 4XR 60 Appendix US ...
Page 61: ...ALTAIR 4XR 61 Appendix US 7 8 Alarm Setup ...
Page 62: ...ALTAIR 4XR 62 Appendix US 7 9 Time and Date Setup ...
Page 63: ...sales norrscope com ALTAIR 4XR 63 Appendix US 7 10 Bluetooth Setup ...