ALTAIR
®
4XR
44
Technical Data
US
5
Technical Data
* Certified for 0 - 25 vol.% O
2
The technical and performance specifications for the specialized EX-H and EX-M sensor are the same
as for the standard EX sensor.
Weight
7.9 oz./224 g (device with battery and clip)
Dimensions
(L x W x H)
4.4 x 3.00 x 1.37 inches/112 x 76 x 33 mm
– without fastening clip
Alarms
Four gas alarm LEDs, a charge status LED, an audible alarm and a vibrating
alarm
Volume of
audible alarm
95 dBA at 30 cm typical
Display
LCD display
Battery type
Rechargeable Li polymer battery.
Must not be charged in Ex area.
Device run time
(Bluetooth OFF)
24 hours at 25 °C (77 °F)
Charging time
4 hours
The maximum safe area charging voltage U
m
= 6.7 VDC
Warm up time
2 min
Temperature range
14 to 104 °F (-10 to 40 °C) Normal operating range
-40 to 140 °F (-40 to 60 °C) Extended operating range
50 to 95 °F (10 to 35 °C) While charging battery
-40 to +140 °F (-40 to 60 °C) Intrinsic safety
ambient temperature range (ATEX, IEC)
-40 to +129 °F (-40 to 54 °C) Intrinsic safety ambient temperature range
(CSA, ETL)
Humidity range
15 % – 90 % relative humidity, non-condensing,
5 % – 95 % RH intermittent
Atmospheric pres-
sure range
800 to 1200 mbar
Ingress
protection
IP 68 (2 meters for 45 minutes)
Measuring
methods
Combustible gases: Catalytic sensor
Oxygen: Electrochemical sensor
Toxic gases: Electrochemical sensor
Combustible
O
2
*
CO
H
2
S
Measuring
range
0-100% LEL
0-5.00% Vol.CH
4
0-30% Vol.
0-1999 ppm
0-1999 mg/m
3
0-200 ppm
0-284 mg/m³
H
2
S-LC
NO
2
CO-H
2
SO
2
0-100 ppm
0-50 ppm
0-1999 ppm
0-20 ppm
Summary of Contents for ALTAIR 4XR
Page 1: ...MSAsafety com Operating Manual ALTAIR 4XR Multigas Detector Order No 10175896 00 ...
Page 52: ...ALTAIR 4XR 52 Appendix US 7 Appendix 7 1 Start Up Sequence Power on ...
Page 53: ...ALTAIR 4XR 53 Appendix US 7 2 Fresh Air Setup FAS ...
Page 55: ...ALTAIR 4XR 55 Appendix US ...
Page 56: ...ALTAIR 4XR 56 Appendix US 7 4 Bump Test ...
Page 57: ...ALTAIR 4XR 57 Appendix US 7 5 Options Setup ...
Page 58: ...ALTAIR 4XR 58 Appendix US 7 6 Sensor Setup ...
Page 59: ...ALTAIR 4XR 59 Appendix US 7 7 Calibrations ...
Page 60: ...ALTAIR 4XR 60 Appendix US ...
Page 61: ...ALTAIR 4XR 61 Appendix US 7 8 Alarm Setup ...
Page 62: ...ALTAIR 4XR 62 Appendix US 7 9 Time and Date Setup ...
Page 63: ...sales norrscope com ALTAIR 4XR 63 Appendix US 7 10 Bluetooth Setup ...