background image

OPERATION 

MSA  

10

AutoMaXX - series 

3.

Operation

3.1.  Preparing the device 

Warning! 

The lung governed demand valve may only be put into use in a fully 

maintained and tested condition. If malfunctions or defects are noticed 

prior to use, do not use the lung governed demand valve under any 

circumstances. 
Get the apparatus checked and repaired by an authorised service 

centre. 

Switching off the device (only Positive pressure lung governed demand 

valve) 

Before connecting the device to medium pressure, in order to prevent accidental 

air delivery, push the red operating button to switch off the AutoMaXX. 

For AutoMaXX-AS / AutoMaXX-AS-C: 

If a handwheel / connection adapter is provided, ensure that it is 

screwed into the mask so that the AutoMaXX can be plugged in. 

3.2.  Connecting the AutoMaXX 

(1) Insert plug-in nipple of AutoMaXX into the coupling of the medium pressure

line until it distinctly snaps into place.

(2) Don full face mask and perform tightness check (palm test)

(

Mask operating manual).

(3) Open cylinder valve(s) fully.
(4) Screw AutoMaXX into the mask (AutoMaXX-AE, AutoMaXX-N) or insert it

into the adapter of the mask until it distinctly snaps into place

(AutoMaXX-AS).

Support during donning 

The AutoMaXX must be attached to the mask and inspected by a second person. 

3.3.  Putting the AutoMaXX into operation (only Positive pressure lung 

governed demand valve) 

To activate, inhale once forcefully or push flush button. 

Warning! 

When using the flush button or bypass, the compressed air 

consumption increases. 

Summary of Contents for AutoMaXX Series

Page 1: ...MSAsafety com Operating Manual AutoMaXX Series Negative and Positive Pressure Lung Governed Demand Valve Order No D4085399 05...

Page 2: ...MSA Europe GmbH Schl sselstrasse 12 8645 Rapperswil Jona Switzerland MSA 2018 All rights reserved...

Page 3: ...governed demand valve 11 3 5 Removing the AutoMaXX 11 3 6 Bypass system only for AutoMaXX AS 11 3 7 Additional function 11 4 Care and maintenance 12 4 1 Maintenance instructions 12 4 2 Maintenance int...

Page 4: ...toMaXX series 5 Technical data 17 6 Approved system components 18 7 Ordering information 20 7 1 Positive pressure lung governed demand valves 20 7 2 Negative pressure lung governed demand valves 21 7...

Page 5: ...taken into account for a safe use Alternative use or use outside these specifications will be considered as non compliance This also applies especially to unauthorised alterations to the apparatus and...

Page 6: ...pass AutoMaXX AS C Lung Governed Demand Valve plug in threaded connection M45X3 Silicon line 650 Combination plug in M45x3 AutoMaXX AS C G Lung Governed Demand Valve plug in threaded connection M45X3...

Page 7: ...ction M45x3 according to DIN EN 148 3 suitable for full face masks PF AS C Positive pressure offers the options AS and AE as combination suitable for full face masks PF N Negative pressure round threa...

Page 8: ...pressure lung governed demand valve Example AutoMaXX AE 1 Handwheel only AE and AS C 9 Diaphragm fork 2 Red operating button 10 Marking Year of manufacturing 3 Housing 11 Flush button 4 Diaphragm 12 P...

Page 9: ...g governed demand valve AutoMaXX N 1 Handwheel 8 Diaphragm groove 2 Operating button 9 Diaphragm fork 3 Housing 10 Marking Year of manufacturing 4 Diaphragm 11 Flush button 5 Protective cap 12 Diaphra...

Page 10: ...ection adapter is provided ensure that it is screwed into the mask so that the AutoMaXX can be plugged in 3 2 Connecting the AutoMaXX 1 Insert plug in nipple of AutoMaXX into the coupling of the mediu...

Page 11: ...ously and separate AutoMaXX from the mask Alternative for AutoMaXX AS C Unscrew handwheel 2 Remove medium pressure line Push the plug in nipple into the coupling of the medium pressure line forcefully...

Page 12: ...rit or petrol 4 2 Maintenance intervals MSA recommends the following maintenance intervals If needed and by considering the usage tasks may be at even shorter intervals than indicated Observe national...

Page 13: ...ith lots of clean water After cleaning disinfection 1 Connect switched off positive pressure AutoMaXX to the medium pressure 2 Push flush button fully blow out the moisture for at least 20 s 3 Uncoupl...

Page 14: ...g buttons with the other hand push together both snap on hooks and push off the protective cap For AutoMaXX N Turn handwheel until the safety clamp below an operating button is visible through the han...

Page 15: ...andwheel until stopper still holding down both operating buttons 5 Perform tightness check and check closing pressure positive pressure versions or rather opening pressure negative pressure versions 4...

Page 16: ...tly or via test adapter line Close all lines of the test equipment Warning Use a test adapter line to prevent damage to test equipment not protected against positive pressure Close the connection betw...

Page 17: ...ons the AutoMaXX must be inspected by MSA or trained service personnel 5 Technical data Operating pressure 5 8 bar dynamic Operating pressure for MicroMaXX series 5 9 bar dynamic Operating temperature...

Page 18: ...AutoMaXX AS C G AutoMaXX AS C short and AutoMaXX AS C G short Compressed air breathing apparatus Series Full face mask Series BD 96 3S PF AirMaXX Ultra Elite PF BD Compact Z 3S H PF F1 Ultra Elite H P...

Page 19: ...XX N short and AutoMaXX N G short Compressed air breathing apparatus Series Full face mask Series BD 96 3S BD AirMaXX Ultra Elite BD Compact Z 3S H F1 Ultra Elite H F1 AutoMaXX N MicroMaXX Compressed...

Page 20: ...B with Bypass Plug in connection Silicon medium pressure line 10038296 AutoMaXX AS B G with Bypass Plug in connection Rubber medium pressure line 10038410 AutoMaXX AS C Plug in connection M45x3 Silico...

Page 21: ...Accessories Description Part No Diaphragm pack of 5 replacement 10032083 Seal cap protective sleeve for plug in nipple pack of 10 D4080871 Seal cap for MicroMaXX series protective sleeve for plug in n...

Page 22: ...For local MSA contacts please visit us at MSAsafety com Because every life has a purpose...

Reviews: