background image

1. Prepare a cleaning solution by adding Confidence

Plus

®

Cleaning Solution (P/N 10009971) to water, fol-

lowing the instructions on the Confidence Plus
Cleaning Solution container.

2. Cleaning by wiping with a damp sponge or cloth con-

taining Confidence Plus Cleaning Solution from MSA.
Follow the directions on the Confidence Plus Cleaning
Solution container for mixing directions and recom-
mended times.

CLEANING AND MAINTENANCE NIGHTFIGHTER
HEADS-UP DISPLAY SYSTEM

The NightFighter Heads-Up Display System should be
cleaned and disinfected after each use. Follow an estab-
lished cleaning and disinfecting program. Failure to follow
this procedure can damage the Nightfighter Heads-Up
Display System.

Remove the Receiver from the Apparatus

Unthread the thumbscrew of NigthFighter Heads-Up
Display System Receiver and slide the receiver from the
facepiece bracket.

INSPECTION OF THE NIGHTFIGHTER HEADS-UP DIS-
PLAY SYSTEM RECEIVER

1. Inspect the receiver module. Look for cracks or other

signs of damage, which could allow contaminants to
enter the module housing. Check that the battery
compartment is clear of moisture or debris. Also
check that the battery compartment Gasket is free of
debris and not damaged or missing.

2. Reassemble the receiver module on the Ultra Elite

Facepiece.

3. Function Check

Note:

If a low battery is present in the receiver, ICM Tx

Unit, or Nightfighter Heads-Up Display System
Transmitter, the LED Yellow light will flash after going
through the start-up sequence, (see “Low Battery
Warnings” in the “During Use” section of this manual).

INSPECTION OF THE ICM TX UNIT

Inspect the entire SCBA and the ICM Tx Unit after EACH
use and MONTHLY.

Note:

Most performance properties of the device cannot

be tested by the user in the field.

1. Perform the ICM Tx Unit Functional Test (outlined in

the “Functional Check of the ICM Tx Unit” portion of
the “Using the ICM Tx Unit” section of this manual.

2. If all the steps are performed successfully, remove the

SCBA and inspect the ICM Tx Unit as follows:

a. Check for external cracks in the case or housing.
b. Check for missing screws or loose covers.
c. Check for signs of leaking covers or water retention

in the case.

d. Check rubber cover for damage.
e. Check for damaged or missing buttons on the unit.
f. Check for any visible signs of damage to compo-

nents.

g. Check for proper operation of buttons. The buttons

should be free from damage or cracks.

h. Check for proper and secure attachment of all com-

ponents to the hose and gauge. If the SCBA passes
all the inspections specified in the SCBA instruction
manual, and the ICM Tx Unit passes all inspections
specified above, the unit is ready to be placed in
service.

i. Check for low battery.

Note:

If the battery in either the ICM Tx Unit or the

NightFighter Heads-Up Display System Receiver is low,
the Yellow LED will flash in the receiver display after going
through the start-up sequence. (see “Low Battery
Warnings” in the “During Use” section of this manual)

Note:

If the SCBA and ICM Tx Unit do not meet all

inspection requirements, the unit must be removed from
service until those conditions which failed are corrected
and the SCBA is re-inspected satisfactorily.

All repair and replacement of subassemblies must be
carried out by a trained repair person. Do not disas-
semble the ICM Tx Unit beyond the limits of this
instruction manual. Failure to follow this warning will
void the NIOSH certifications and can result in serious
personal injury or death.

STORAGE AND MAINTENANCE

1. In order to maintain the ICM Tx Unit and NightFighter

Heads-Up Display System Receiver, they should be
cleaned after each use with a damp cloth using soap
and water. DO NOT APPLY CLEANING SOLVENTS.

2. Units contaminated by chemical or radioactive materi-

als must be decontaminated or disposed of according
to all applicable regulatory standards.

3. In the event of any malfunction, return the unit for

repair. Contact MSA Customer Service for return
authorization.

4. Store the units in a cool, dry, ventilated area at ambi-

ent temperatures consistent with the battery manufac-
turer’s recommendations. To store the NightFighter
Heads-Up Display System components, be sure that
the ICM Tx Unit is in the OFF (LED is not illuminated).
For prolonged storage, remove the batteries to pre-
vent battery corrosion.

CLEANING AND MAINTENANCE

15

TAL 709 (L) Rev. 4 - 10058881

Summary of Contents for ICM Tx

Page 1: ...Display System which integrates with the SCBA These instructions in addition to the instructions supplied with the SCBA must be carefully read and followed by all persons who use or maintain this prod...

Page 2: ...epiece Note The NightFighter Heads Up Display can only be used with an Ultra Elite Facepiece 2 The NightFighter Heads Up Display consists of three 3 separate assemblies Bracket assembly attached to an...

Page 3: ...nation of batteries from dif ferent manufacturers will affect the performance of unit and void the Intrinsic Safety Approval Intrinsically Safe Rating The NightFighter Heads Up Display System and ICM...

Page 4: ...4 TAL 709 L Rev 4 10058881 NOTES...

Page 5: ...he facepiece which displays the air cylinder pressure in one quarter cylinder increments by an LED light pattern See chart 1 Assembly of Receiver to Facepiece Bracket With the facepiece laying on its...

Page 6: ...6 TAL 709 L Rev 4 10058881 NOTES...

Page 7: ...m the unit o The center lighted ALARM button green red acti vates the full alarm mode with or without air pres sure o The top MODE button green will refresh the receiv er display set the receiver to c...

Page 8: ...8 TAL 709 L Rev 4 10058881 NOTES...

Page 9: ...Pressure Full Alarm with or without Pressure Deactivation of Full Alarm Deactivation of PreAlarm with Shake or move Unit Low Battery Thermal Alarm Activation see Thermal Alarm Activation Curve Redunda...

Page 10: ...trinsic Safety Approval is void Failure to follow this warning can result in serious personal injury or death Before using the ID tag inspect for damage or cracks in the case If damage is found discar...

Page 11: ...he bypass valve on the second stage regulator to bleed off pressure until the ICM Tx Unit pressure reading drops below the following values a 530psig approximately 2216psi system b 750psig approximate...

Page 12: ...NOTES 12 TAL 709 L Rev 4 10058881...

Page 13: ...injury or death Actual time remaining may be less than the calculated time displayed Increases in breathing rate may reduce remaining time more than expected Use time indica tor as general guide only...

Page 14: ...sure to tighten the battery cover screws enough to provide adequate compression on the battery cover gasket to prevent dirt and moisture from entering the battery compart ment A torque of 6 in lbs is...

Page 15: ...for missing screws or loose covers c Check for signs of leaking covers or water retention in the case d Check rubber cover for damage e Check for damaged or missing buttons on the unit f Check for any...

Page 16: ...cover 4 Before replacing the battery cover be sure that the rubber cover on the ICM Tx Unit is properly engaged with the plastic housing The rubber cover has barbs that hold it firmly against the plas...

Page 17: ...transfilling hose assembly or fittings other than those supplied by MSA specifically for the Quick Fill System Use of any other transfilling hose assembly and or fitting may result in serious personal...

Page 18: ...System hose to the SCBA c Remove the rubber dust cap from the male inlet fit ting on the SCBA Be sure that the cylinder valve is fully opened d Remove the rubber dust cap from the female fitting on th...

Page 19: ...he male Quick Fill System fitting d Push the female fittings on to the male fittings until they click in place Pull on the hose to be sure it snapped in place If serious leakage is noticed from either...

Page 20: ...NOTES 20 TAL 709 L Rev 4 10058881...

Page 21: ...emblies 485331 802687 802688 802689 802690 and 48332 for filling cylinders in IDLH atmospheres DO NOT Transfill be a Donor using a 3000psi URC Assembly The 3000psi URC Assembly has a check valve that...

Page 22: ...either female fitting or along the hose depressurize the hose and correct the prob lem Such leakage can result in increased fill time 2 To attach the Quick Fill System hose to the URC Assembly a Remo...

Page 23: ...c Quickly replace the protective dust cap on the URC Assembly male regulator fitting This will form a redundant seal d Open the cylinder valve and return to fresh air immediately The dust cover preven...

Page 24: ...eration and instructions manual PN 10023638 for additional war ranty information For a copy of the complete warranty or for information on submitting a warranty claim write to MSA Customer Service Dep...

Reviews: