3.8
Installing the Breathing Tube
3.8.1 Attaching the Breathing Tube Assembly to the Blower
(1)
Inspect the o-ring in the breathing tube inlet. If the o-ring is damaged, missing, or the
breathing tube is difficult to install, replace the o-ring. Reorder PN 10085084 (Pack
of 10).
(2)
Attach the breathing tube inlet to the blower outlet connection by aligning the marks
on the bayonet fitting and turning the breathing tube to align the arrow on the unit to
the second mark on the breathing tube.
(3)
Ensure that the bayonet flange is positioned properly.
3.8.2 Attaching the Breathing Tube Assembly to the Full Coverage Hoods
There are two methods (locking clamp, threaded connector) of attaching the breathing tube assembly to Full Coverage material
hoods. Breathing tube assembly (P/N 10049631 and 10082281) may be used with the hoods. P/N 10082281 is approximately
4” longer than P/N 10049631.
NOTE:
Ensure the hood and components are not torn or damaged.
OptimAir TL PAPR
19
US
3 Preparing the OptimAir TL PAPR for Use