SingleLine SCOUT
10
Use
GB
Alarm Statuses
The user can acknowledge an alarm after it is triggered.
Cylinder Pressure Display
The cylinder pressure is displayed via the integrated analogue gauge and additionally detected from
an internal pressure sensor. Current pressure can be read from the gauge using the imprinted scale.
The up-to-date filling level of the compressed air cylinders is displayed in 3 stages, in accordance with
the following coding, using 3 LEDs on each side of the gauge when the device is lifted from the user’s
chest.
Following thresholds represent the standard setting and can be changed using MSA A2 Software in
combination with 10181672 as PC interface.
See MSA A2 Software for more details.
Alarm
LEDs
Acoustic
1st threshold
Gauge green lighted and blinking
1 Beep
2nd threshold
Gauge yellow lighted and blinking
2 Beep
60 bar
Gauge red & buddy lights lighted
and blinking
2 Beep
< 60 bar
Gauge red & buddy lights lighted
and blinking
2 Beep
Battery warning
Bottom red
None
Manual alarm
Gauge red & buddy lights lighted
and blinking
Distinct pattern
Motion pre-alarm
Gauge red & buddy lights lighted
and blinking
Increasingly louder consecutive
variable tones
Motion alarm
Gauge red & buddy lights lighted
and blinking
Distinct pattern
Alarm
Acknowledge
1st threshold
Not necessary, since short acoustic signal.
2nd threshold
Not necessary, since short acoustic signal.
60 bar
Confirmation only possible below 15 bar pressure: Push the alarm
button twice.
Pneumatic whistle cannot be confirmed.
< 60 bar
Alarm will continuously appear.
Battery warning
Exchange batteries – outside of hazardous area only
Replacing the battery after the operation.
Manual alarm
Push the alarm button twice.
Motion pre-alarm
Movement of SingleLine SCOUT
Motion alarm
Push the alarm button twice.