44
Operation
Gas Table
The ULTIMA XIR PLUS combustible sensor can be calibrated to a wide range of compounds, see
section 9 for a list of gases, span values, and gas table values.
The XIR PLUS sensor is only performance approved for Methane and Propane.
Gas Table #
Target Gas
1
Methane
2
Propane
3
Ethane
4
Butane
5
Pentane
6
Hexane
7
Cyclopentane
8
Ethylene
Note that the Gas Table numbers listed above represent different linearity curves for infrared
absorbing gases. Typically, higher Gas Table numbers represent weaker absorbing gases. Each
curve represents a linearity that is applicable to a group of hydrocarbon gases and not necessarily
one specific gas. The Target Gas shown is representative of a group of gases with similar linearity.
WARNING!
The XIR PLUS sensor must be calibrated after changing the gas table and/or span value.
Failure to follow the above warning can result in serious personal injury or loss of life.
Reset Sensor
The sensor default values can be restored by resetting the sensor. If Swap Delay is enabled,
during a sensor reset the analog outputs will behave the same as if a sensor were replaced. See
the section on Swap Delay for details. If Swap Delay is disabled the analog outputs will first go
into a fault condition, followed by the maintenance condition for the duration of the sensor’s count-
down.
NOTICE
The sensor goes into a Sensor Configuration Reset fault and must be calibrated after a sensor
reset. There will be four dashes displayed where the current reading should be.
All settings, including Alarm Set Point and Calibration Values, will be returned to factory defaults.
To Reset Sensor to factory default:
(1) Scroll to Settings and select it.
(2) Select Sensor.
(3) Scroll and select Reset Sensor.
(4) Select Continue.
Note: Sensor Reset can take 10 seconds to initiate the sensor’s countdown.
(5) Calibrate the sensor to clear Sensor Configuration Reset fault.
US
Summary of Contents for TG5000
Page 1: ...Operating Manual TG5000 Gas Monitor Order No 10212126 L Y REV 0 US ...
Page 36: ...36 Operation US ...
Page 46: ...46 Operation US ...
Page 62: ...62 Maintenance US ...
Page 63: ...63 Maintenance US ...
Page 64: ...64 Maintenance US ...
Page 70: ...70 Appendix Calibration Guide and Additional Gases US ...
Page 71: ...71 Appendix Calibration Guide and Additional Gases R 32 US ...
Page 72: ...72 Appendix Calibration Guide and Additional Gases US ...
Page 75: ...For local MSA contacts please visit us at MSAsafety com ...