40
6.
Place the demonstration plate in the single plate adapter with the chamfered corners of the plate facing into the instrument
(
Figure 5:7 MESO SECTOR demonstration plate in the single plate adapter
NOTE: If the SECTOR S 600 is unable to read the barcode on the demonstration plate (or if no barcode label is present), the
plate will be skipped and ejected from the instrument. Please contact MSD
for assistance.
5.3
Run
When the camera reaches proper operating temperature, the Instrument Log (click Show Log in the Status window) will indicate that the
temperature is locked. At that point, the instrument is ready.
Click Run in the SECTOR S 600 window. The Run Options dialog window will open (
1.
Verify Setup Selections. If changes are necessary, click Cancel, make changes in the area (
) of the SECTOR S 600
window, and re-verify Setup Selections.
2.
Run name is optional, but one can be entered in the second section.
3.
Verify the Export information and make changes if necessary. (Refer to the DISCOVERY WORKBENCH User’s Guide for
additional details.) The export selections control the format and the location of the exported text data file that will be created
when the demonstration plate is read.
4.
Click
OK
. The plate read starts. View the status in the lower left region (Status bar) of the DISCOVERY WORKBENCH software
to monitor progress. When the plate read is complete, the demonstration plate is returned to the stacker and
Run
is enabled
again.
Summary of Contents for MESO SECTOR S 600
Page 1: ...MESO SECTOR S 600 www mesoscale com USER MANUAL ...
Page 6: ...6 1 Introduction ...
Page 9: ...9 2 SafetyInformation ...
Page 19: ...19 3 SystemDescription ...
Page 33: ...33 4 Installation ...
Page 35: ...35 5 QuickStart ...
Page 43: ...43 6 UsingtheSECTORS600 ...
Page 47: ...47 7 Maintenance ...
Page 50: ...50 8 Appendix ...
Page 56: ...56 9 TechnicalSupport ...