49
7.2.1
Adapter and Stack Tube Cleaning
Spills on single plate adapters or stack tubes should be promptly cleaned to prevent a build-up of crystalline debris or dried salts that
could interfere with their operation. In particular, plates may not load correctly if the latch movement of these devices is compromised.
Clean dirty latch mechanisms by soaking the latches in warm water (approximately 104°F/40°C) with a small amount of residue-free
detergent for 15 minutes. Rinse the latches thoroughly with running water, tip and/or shake gently to remove excess liquid, and allow to
fully air dry. If there is any residue remaining, soak the latches in a 70% alcohol solution (isopropanol or ethanol in water) for 15 minutes;
shake out the excess, and allow to air dry completely. If plate-loading problems persist, contact
7.3
Instrument Decontamination
Contact MSD
for a detailed protocol for instrument decontamination.
NOTE: The SECTOR S 600 must be decontaminated prior to shipping the instrument back to MSD. Contact MSD
prior to shipping for instructions and requirements.