Using the SECTOR S 600
This chapter explains how to prepare and load single or multiple MULTI-ARRAY plates for typical use with the SECTOR S 600. Additional
information is provided to integrate the instrument with an external, third-party robotic system. Please refer to the DISCOVERY
WORKBENCH User’s Guide for detailed instructions on running MULTI-ARRAY plates on the SECTOR S 600 system.
NOTE: This section assumes that a qualified MSD Service Engineer has installed and configured the instrument, and that the user
has run a demonstration plate as described in the previous section.
6.1
Single Plate Run
To run a single MULTI-ARRAY plate, place the single plate adapter in the (right) stacker interface.
The right stacker interface can act as both input and output locations. If not using the output (left) interface, place the stacker cover plate
over it. The SECTOR S 600 will pull and eject plates from/to the input (right) interface.
If using the output side, install a second single plate adapter or a stack tube on the output (left) stacker interface plate. The SECTOR S
600 will automatically pull plates from the right side and eject plates into the left side.
If the stacker cover plate is present on the output stack, remember to select the Return Plates to Input Stack
checkbox so that the SECTOR S 600 will not attempt to place the plate into the covered stack at the end of the
read.
Prepare the MULTI-ARRAY plate using the required reagents per the assay protocol. Place the plate into the single plate adapter. Sensors
within the instrument will automatically identify the plate orientation and compensate the data acquisition and analysis accordingly. Per
the detailed instructions in the DISCOVERY WORKBENCH User’s Guide, configure the options in DISCOVERY WORKBENCH and run
the plate.
6.2
Stack Run—Multiple Plates
Multiple plates may be loaded into SECTOR S 600 instrument using the standard or high-capacity stack tubes (
). SECTOR S
600 reads plates from the bottom of the stack tube.
6.2.1
Loading the Stack Tube
Prepare MULTI-ARRAY plates using the required reagents per the assay protocol.
To load prepared plates into the stack tube:
1.
Stack plates on a stable surface.
2.
Place the stack of plates on top of an extra, standard 96-well MULTI-ARRAY plate. This extra plate serves as a spacer for
loading (
3.
Orient the stack tube so that the corners of the plates are toward the back of the instrument.
Summary of Contents for MESO SECTOR S 600
Page 1: ...MESO SECTOR S 600 www mesoscale com USER MANUAL ...
Page 6: ...6 1 Introduction ...
Page 9: ...9 2 SafetyInformation ...
Page 19: ...19 3 SystemDescription ...
Page 33: ...33 4 Installation ...
Page 35: ...35 5 QuickStart ...
Page 43: ...43 6 UsingtheSECTORS600 ...
Page 47: ...47 7 Maintenance ...
Page 50: ...50 8 Appendix ...
Page 56: ...56 9 TechnicalSupport ...