Appendix
54
8.2
Specifications
8.2.1
SECTOR S 600
NOTE: All performance information including standard signal levels, detection limits, dynamic range, and instrument noise
is based on 150
µ
L read volumes in 96-well MULTI-ARRAY plates using standard plate read parameters.
8.2.2
Scientific Performance
Table 8:2 Scientific performance
SECTOR S 600
Standard Signal Levels
Signals generated by Free TAG, 15,000 [Cat # R31QQ-3]
Signals generated by PQ Low Control [Cat # R31QQ-3]
12,000–18,000 counts
25–100 counts
Dynamic Range
10
6
Electronic Noise
Standard Deviation Dark Noise (one spot plates)
≤
16 counts
8.2.3
Environmental Specifications
Operational Temperature Range
20–26
°
C (68–78
°
F)
Operational Humidity Range
10–80% non-condensing
Storage Temperature Range
5–40
°
C (41–104
°
F)
Storage Humidity Range
5–85% non-condensing
Ambient Light
< 2000 Lux
8.2.4
Power Requirements
Voltage
100–240 V~ 50/60 Hz
Current
3.5A–2.8A @ 100–240 V~
Class I, Installation (Overvoltage) Category II Device
Fuses
Electrical fuses are only accessible by MSD Service Engineers and are not user-serviceable. The information below is provided for
reference.
Main Fuses F1 and F2:
T4.0 A, 250 V, 100–240 V~
Summary of Contents for MESO SECTOR S 600
Page 1: ...MESO SECTOR S 600 www mesoscale com USER MANUAL ...
Page 6: ...6 1 Introduction ...
Page 9: ...9 2 SafetyInformation ...
Page 19: ...19 3 SystemDescription ...
Page 33: ...33 4 Installation ...
Page 35: ...35 5 QuickStart ...
Page 43: ...43 6 UsingtheSECTORS600 ...
Page 47: ...47 7 Maintenance ...
Page 50: ...50 8 Appendix ...
Page 56: ...56 9 TechnicalSupport ...