21
one side
click sound card
Incorrect installation of
stage
Re install the stage
Incorrect installation of
objective nosepiece
Push the dovetail joint all the way
to the inside
The condenser is out of
alignment
Re-centering the condenser
The thickness of the
sample is not uniform
Replace the sample or prepare the
sample again
The image
fluctuates
Common
Incorrect installation of
objective turntable
Push the dovetail joint all the way
to the inside
The objective nosepiece
is not in place
Make sure that the objective
nosepiece is in the position of the
click sound card
The condenser is out of
alignment
Re-centering the condenser
The focus
knob is
difficult to
turn
Common
The adjustment of the
tightening device of the
focusing knob is too tight
Loosen the adjusting knob elastic
device
After locking the limit
device of coarse
collimation knob, try to
adjust the stage with
Turn on the limit device of coarse
quasi focus knob