22
coarse collimation knob
Image cannot
be focused /
the stage
slides
automatically
/ the coarse
and fine focus
knob slides
Common/
fluorescen
ce
Too loose adjustment of
coarse collimation knob
tightening device
Tighten the loose device of coarse
collimation knob
The coarse
focusing can
not rise all
the time
Common/
fluorescen
ce
The coarse collimation
knob limit device blocks
the stage
Turn on the limit device of coarse
focusing knob
Coarse focus
can never be
reduced
Common/
fluorescen
ce
The condenser frame is
too low
Raise the condenser frame
Before
focusing, the
objective
collides with
the sample
Common/
fluorescen
ce
The sample is upside
down
Install the sample correctly
The field of
view of one
Common
The pupil distance is not
adjusted correctly
Correct adjustment of pupil
distance